Performance tests using the Lexsyg luminescence reader

This study presents test measurements on a Lexsyg luminescence reader, acquired by the Giessen luminescence group. The reader is of the type Standard, hence designed for routine determination of palaeodoses using quartz or K-feldspars. The tests include measurements of the stimulation powers, prehea...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Geochronometria 2014-12, Vol.41 (4), p.327-333
Hauptverfasser: Lomax, Johanna, Kreutzer, Sebastian, Fuchs, Markus
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:This study presents test measurements on a Lexsyg luminescence reader, acquired by the Giessen luminescence group. The reader is of the type Standard, hence designed for routine determination of palaeodoses using quartz or K-feldspars. The tests include measurements of the stimulation powers, preheat temperatures, OSL- and TL-curves, as well as dose recovery tests and calibration measurements using highly sensitised calibration quartz. A comparison of D e values determined using a Lexsyg reader with single grain D e -values, which have been previously obtained on a Risø reader, is also presented. The results imply that the Lexsyg reader is a highly reliable measurement device with high reproducibility, yielding instrumental uncertainties of around 0.3%.
ISSN:1733-8387
1897-1695
DOI:10.2478/s13386-013-0174-x