Performance of the EDELWEISS-III experiment for direct dark matter searches

We present the results of measurements demonstrating the efficiency of the EDELWEISS-III array of cryogenic germanium detectors for direct dark matter searches. The experimental setup and the FID (Fully Inter-Digitized) detector array is described, as well as the efficiency of the double measurement...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of instrumentation 2017-08, Vol.12 (8), p.P08010-P08010
Hauptverfasser: Armengaud, E., Arnaud, Q., Augier, C., Benoît, A., Bergé, L., Bergmann, T., Billard, J., de Boissière, T., Bres, G., Broniatowski, A., Brudanin, V., Camus, P., Cazes, A., Chapellier, M., Charlieux, F., Jésus, M. De, Dumoulin, L., Eitel, K., Filosofov, D., Foerster, N., Fourches, N., Garde, G., Gascon, J., Giuliani, A., Grollier, M., Gros, M., Hehn, L., Hervé, S., Heuermann, G., Humbert, V., Jin, Y., Juillard, A., Kéfélian, C., Kleifges, M., Kozlov, V., Kraus, H., Kudryavtsev, V. A., Le-Sueur, H., Lin, J., Maisonobe, R., Mancuso, M., Marnieros, S., Menshikov, A., Navick, X.-F., Nones, C., Olivieri, E., Pari, P., Paul, B., Poda, D., Queguiner, E., Robinson, M., Rodenas, H., Rozov, S., Sanglard, V., Schmidt, B., Scorza, S., Siebenborn, B., Tcherniakhovski, D., Vagneron, L., Weber, M., Yakushev, E., Zhang, X., Zolotarova, A.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We present the results of measurements demonstrating the efficiency of the EDELWEISS-III array of cryogenic germanium detectors for direct dark matter searches. The experimental setup and the FID (Fully Inter-Digitized) detector array is described, as well as the efficiency of the double measurement of heat and ionization signals in background rejection. For the whole set of 24 FID detectors used for coincidence studies, the baseline resolutions for the fiducial ionization energy are mainly below 0.7 keVee (FHWM) whereas the baseline resolutions for heat energies are mainly below 1.5 keVee (FWHM). The response to nuclear recoils as well as the very good discrimination capability of the FID design has been measured with an AmBe source. The surface β- and α-decay rejection power of Rsurf
ISSN:1748-0221
1748-0221