Ring wire anemometer—Part II: Measurement of small‐scale turbulence

The ring wire, a circular hot‐wire anemometer probe, orientated normal to the flow, is useful to measure turbulent scales in the dissipative range. It is less sensitive to angular positioning errors than straight wire sensors. Following the rationale of Wyngaard [J. Phys. E. 52, 1, 1105 (1968)], the...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Review of scientific instruments 1987-11, Vol.58 (11), p.2175-2180
Hauptverfasser: Mestayer, Patrice, Giovanangeli, Jean‐Paul, Chambaud, Pierre
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The ring wire, a circular hot‐wire anemometer probe, orientated normal to the flow, is useful to measure turbulent scales in the dissipative range. It is less sensitive to angular positioning errors than straight wire sensors. Following the rationale of Wyngaard [J. Phys. E. 52, 1, 1105 (1968)], theoretical expressions are derived for the high wave number attenuation due to spatial resolution (line averaging). The expressions are applied to the estimation of longitudinal velocity spectrum and dissipation rate. The ring wire is shown to present much smaller attenuations than the straight wire of the same length or, inversely, to have a much higher sensitivity and frequency response than the straight wire of similar spatial resolution. The theoretical results are confirmed by experimental data.
ISSN:0034-6748
1089-7623
DOI:10.1063/1.1139482