Optimization of Vertical Acceptance of the Static Mass Analyzer

Axial aberration of the ion-optical system of a static mass analyzer is one of the factors limiting the transmittance and sensitivity of the device because of a need in significantly decreasing the vertical component of the ion beam emittance to get the required mass spectral resolution. We describe...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of analytical chemistry (New York, N.Y.) N.Y.), 2020-12, Vol.75 (13), p.1660-1664
Hauptverfasser: Sachenko, V. D., Antonov, A. S., Gall, L. N., Berdnikov, A. S.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Axial aberration of the ion-optical system of a static mass analyzer is one of the factors limiting the transmittance and sensitivity of the device because of a need in significantly decreasing the vertical component of the ion beam emittance to get the required mass spectral resolution. We described an algorithm for selecting the optimal collimating system of the mass analyzer, which effectively decreases the contribution of axial aberration with minimal loss of the intensity of mass spectral lines.