Experimental observation and numerical simulation of spectra of solid-anode X-ray tubes
Possibilities of the experimental measurement and theoretical modeling of a fluorescence spectrum of a solid-anode X-ray tube are studied. A setup for the direct measurement of spectra of X-ray tubes of different types is created, spectra of a commercial BKh15 Svetlana X-ray tube are measured in wid...
Gespeichert in:
Veröffentlicht in: | Journal of analytical chemistry (New York, N.Y.) N.Y.), 2016-05, Vol.71 (5), p.471-475 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Possibilities of the experimental measurement and theoretical modeling of a fluorescence spectrum of a solid-anode X-ray tube are studied. A setup for the direct measurement of spectra of X-ray tubes of different types is created, spectra of a commercial BKh15 Svetlana X-ray tube are measured in wide range of anode voltages and currents. Calculations of spectra using the Ebel algorithm (Horst Ebel, 2006) with the parameters recommended in the original work ensure the satisfactory quality of the reproduction of the measured spectra, but demonstrate some statistically noticeable deviations. The optimization of model parameters ensures the improvement of the quality of simulation. |
---|---|
ISSN: | 1061-9348 1608-3199 |
DOI: | 10.1134/S1061934816030151 |