Low-energy calibration of XENON1T with an internal $$^{{\textbf {37}}}$$Ar source
A low-energy electronic recoil calibration of XENON1T, a dual-phase xenon time projection chamber, with an internal $^{37}$$ 37 Ar source was performed. This calibration source features a 35-day half-life and provides two mono-energetic lines at 2.82 keV and 0.27 keV. The photon yield and electron y...
Gespeichert in:
Veröffentlicht in: | Eur.Phys.J.C 2023-06, Vol.83 (6), Article 542 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A low-energy electronic recoil calibration of XENON1T, a dual-phase xenon time projection chamber, with an internal
$^{37}$$
37
Ar source was performed. This calibration source features a 35-day half-life and provides two mono-energetic lines at 2.82 keV and 0.27 keV. The photon yield and electron yield at 2.82 keV are measured to be (
$$32.3\,\pm \,0.3$$
32.3
±
0.3
) photons/keV and (
$$40.6\,\pm \,0.5$$
40.6
±
0.5
) electrons/keV, respectively, in agreement with other measurements and with NEST predictions. The electron yield at 0.27 keV is also measured and it is (
$$68.0^{+6.3}_{-3.7}$$
68
.
0
-
3.7
+
6.3
) electrons/keV. The
$^{37}$$
37
Ar calibration confirms that the detector is well-understood in the energy region close to the detection threshold, with the 2.82 keV line reconstructed at (
$$2.83\,\pm \,0.02$$
2.83
±
0.02
) keV, which further validates the model used to interpret the low-energy electronic recoil excess previously reported by XENON1T. The ability to efficiently remove argon with cryogenic distillation after the calibration proves that
$^{37}$$
37
Ar can be considered as a regular calibration source for multi-tonne xenon detectors. |
---|---|