Development a robust ionic liquid-based dispersive liquid-liquid microextraction against high concentration of salt combined with flame atomic absorption spectrometry using microsample introduction system for preconcentration and determination of cobalt in water and saline samples
A robust ionic liquid-based method for dispersive liquid-liquid microextraction was developed for the determination of cobalt in saline samples. Its robustness was increased by introducing a common ion of the IL into the sample solution. The concentration of cobalt in the settled phase of the ionic...
Gespeichert in:
Veröffentlicht in: | Mikrochimica acta (1966) 2011-02, Vol.172 (1-2), p.75-82 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A robust ionic liquid-based method for dispersive liquid-liquid microextraction was developed for the determination of cobalt in saline samples. Its robustness was increased by introducing a common ion of the IL into the sample solution. The concentration of cobalt in the settled phase of the ionic liquid was determined by the FAAS using a homemade microsample injection valve. Under the optimized conditions, the calibration graph was linear in the concentration range from 0.4 to 120 μg L
−1
with the detection limit of 0.1 μg L
−1
. Accuracy was checked against the ISO standard method. The method was successfully applied to the determination of cobalt in water and saline samples. |
---|---|
ISSN: | 0026-3672 1436-5073 |
DOI: | 10.1007/s00604-010-0406-1 |