GAS-FILLED IONIZATION CHAMBER AND SECONDARY-EMISSION DETECTOR FOR MEASURING [gamma]-RAY DOSE RATE
The design of a KGK-2 [gamma]-ray dose rate detector is described. The basic characteristics of the detector are presented. The readings of a KGK-2 gas-filled ionization chamber detector and VED-2 secondary-electron emission detector during power startup of BR-K1 and BR-1M reactors are compared. It...
Gespeichert in:
Veröffentlicht in: | Atomic energy (New York, N.Y.) N.Y.), 2020-07, Vol.128 (3), p.166 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The design of a KGK-2 [gamma]-ray dose rate detector is described. The basic characteristics of the detector are presented. The readings of a KGK-2 gas-filled ionization chamber detector and VED-2 secondary-electron emission detector during power startup of BR-K1 and BR-1M reactors are compared. It is shown that by using KGK-2 and VED-2 detectors in combination the [gamma]-ray dose rate can be measured in the interval 1 [mu]Gy/sec to 10 GGy/sec with the linear sections of the regulation curves overlapping to at least four decimal orders. |
---|---|
ISSN: | 1063-4258 1573-8205 |
DOI: | 10.1007/s10512-020-00669-5 |