Calibration and Validation of Dynamic Force Transducers
The urgent need for a procedure to use in validation of dynamic force transducers is considered. A model of the dynamic interaction of the elements of a dynamic force transducer in the course of calibration by the test mass method is presented. Mathematical expressions that describe the calibration...
Gespeichert in:
Veröffentlicht in: | Measurement techniques 2017-10, Vol.60 (7), p.691-695 |
---|---|
1. Verfasser: | |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The urgent need for a procedure to use in validation of dynamic force transducers is considered. A model of the dynamic interaction of the elements of a dynamic force transducer in the course of calibration by the test mass method is presented. Mathematical expressions that describe the calibration process are presented. A schematic diagram of a calibration plant that would be most appropriate for dynamic force transducers is proposed. |
---|---|
ISSN: | 0543-1972 1573-8906 |
DOI: | 10.1007/s11018-017-1255-7 |