Evaluation of four aerobiological sampling methods for the retrieval of aerosolized Pseudomonas syringae

The Andersen six-stage impactor, the SAS (Surface Air System) impactor, the AGI-30 impinger, and gravity plates were evaluated for the retrieval of aerosol-released Pseudomonas syringae. The upper limits of the impactor samplers were exceeded at a spray concentration of 10(7) CFU/ml, indicating that...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Applied and Environmental Microbiology 1991-04, Vol.57 (4), p.1268-1270
Hauptverfasser: Buttner, M.P. (University of Nevada, Las Vegas, Nevada), Stetzenbach, L.D
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The Andersen six-stage impactor, the SAS (Surface Air System) impactor, the AGI-30 impinger, and gravity plates were evaluated for the retrieval of aerosol-released Pseudomonas syringae. The upper limits of the impactor samplers were exceeded at a spray concentration of 10(7) CFU/ml, indicating that these samplers are not appropriate for monitoring high airborne concentrations. Decreased cell concentrations were retrieved with increased sampling time for the Andersen and AGI samplers, indicating that a minimum sampling time is preferable for monitoring aerosolized vegetative cells
ISSN:0099-2240
1098-5336
DOI:10.1128/aem.57.4.1268-1270.1991