MEASUREMENT OF ENERGY FLUX DENSITY DISTRIBUTION IN THE FOCUS OF AN ARC IMAGE FURNACE
A calorimeter for mapping flux density in a focal volume is described. The design is based on an earlier design by Broido and Willoughby. It differs chiefly in two ways. First it is much smaller in order to permit its placement inside a test specimen chamber. Second the design provides for an apertu...
Gespeichert in:
Hauptverfasser: | , , |
---|---|
Format: | Report |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A calorimeter for mapping flux density in a focal volume is described. The design is based on an earlier design by Broido and Willoughby. It differs chiefly in two ways. First it is much smaller in order to permit its placement inside a test specimen chamber. Second the design provides for an aperture-containing shield which is placed in front of the receiving surface calorimeter in order to limit the area of beam which is intercepted. The decay constant is 0.002/sec. Flux densities ranging from 5 to 120 cal/sq cm-sec can be conveniently measured over intercepted beam areas of from 0.03 to 0.2 sq cm. With this calorimeter the flux-density distribution in the focal volume of a double ellipsoidal are image furnace system was determined. (Author)
Pub. in Applied Optics v3 n1 p131-5 Jan 1964 (Copies not available to DDC or Clearinghouse customers). |
---|