THORACIC CAGE IMPEDANCE MEASUREMENTS IMPEDANCE PRODUCT SYSTEM
The measurement of (1) peak-to-peak amplitude of a percent change of impedance measured acrowss the thorax, (2) heart rate as a beat-to-beat d.c. voltage level, and (3) the product of these two variables is performed by an impedance product system. Results show tat the peak amplitude can be measured...
Gespeichert in:
Hauptverfasser: | , |
---|---|
Format: | Report |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The measurement of (1) peak-to-peak amplitude of a percent change of impedance measured acrowss the thorax, (2) heart rate as a beat-to-beat d.c. voltage level, and (3) the product of these two variables is performed by an impedance product system. Results show tat the peak amplitude can be measured by a charge storage and transfer system accurately to within 3%. The cardiac tachometer is linear within 2.5% in a range from 50 to 250 beats/min. TMultiplication of the two variables is accomplished with Pace TR-10 analog computer components with an error of less than 1%. Overall accuracy of the system appears to be within 5%. Tests on several subjects illustrate the change in impedance product with time following the cessation exercise. (Author) |
---|