The Use of Pulse Oximetry in the Assessment of Acclimatization to High Altitude

: Finger pulse oximeters are widely used to monitor physiological responses to high-altitude exposure, the progress of acclimatization, and/or the potential development of high-altitude related diseases. Although there is increasing evidence for its invaluable support at high altitude, some controve...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Sensors (Basel, Switzerland) Switzerland), 2021-02, Vol.21 (4), p.1263
Hauptverfasser: Dünnwald, Tobias, Kienast, Roland, Niederseer, David, Burtscher, Martin
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:: Finger pulse oximeters are widely used to monitor physiological responses to high-altitude exposure, the progress of acclimatization, and/or the potential development of high-altitude related diseases. Although there is increasing evidence for its invaluable support at high altitude, some controversy remains, largely due to differences in individual preconditions, evaluation purposes, measurement methods, the use of different devices, and the lacking ability to interpret data correctly. Therefore, this review is aimed at providing information on the functioning of pulse oximeters, appropriate measurement methods and published time courses of pulse oximetry data (peripheral oxygen saturation, (SpO ) and heart rate (HR), recorded at rest and submaximal exercise during exposure to various altitudes. : The presented findings from the literature review confirm rather large variations of pulse oximetry measures (SpO and HR) during acute exposure and acclimatization to high altitude, related to the varying conditions between studies mentioned above. It turned out that particularly SpO levels decrease with acute altitude/hypoxia exposure and partly recover during acclimatization, with an opposite trend of HR. Moreover, the development of acute mountain sickness (AMS) was consistently associated with lower SpO values compared to individuals free from AMS. : The use of finger pulse oximetry at high altitude is considered as a valuable tool in the evaluation of individual acclimatization to high altitude but also to monitor AMS progression and treatment efficacy.
ISSN:1424-8220
1424-8220
DOI:10.3390/s21041263