Intraoperative Beat-to-Beat Pulse Transit Time (PTT) Monitoring via Non-Invasive Piezoelectric/Piezocapacitive Peripheral Sensors Can Predict Changes in Invasively Acquired Blood Pressure in High-Risk Surgical Patients
Non-invasive tracking of beat-to-beat pulse transit time (PTT) via piezoelectric/piezocapacitive sensors (PES/PCS) may expand perioperative hemodynamic monitoring. This study evaluated the ability for PTT via PES/PCS to correlate with systolic, diastolic, and mean invasive blood pressure (SBP , DBP...
Gespeichert in:
Veröffentlicht in: | Sensors (Basel, Switzerland) Switzerland), 2023-03, Vol.23 (6), p.3304 |
---|---|
Hauptverfasser: | , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Non-invasive tracking of beat-to-beat pulse transit time (PTT) via piezoelectric/piezocapacitive sensors (PES/PCS) may expand perioperative hemodynamic monitoring. This study evaluated the ability for PTT via PES/PCS to correlate with systolic, diastolic, and mean invasive blood pressure (SBP
, DBP
, and MAP
, respectively) and to detect SBP
fluctuations.
PES/PCS and IBP measurements were performed in 20 patients undergoing abdominal, urological, and cardiac surgery. A Pearson's correlation analysis (r) between 1/PTT and IBP was performed. The predictive ability of 1/PTT with changes in SBP
was determined by area under the curve (reported as AUC, sensitivity, specificity).
Significant correlations between 1/PTT and SBP
were found for PES (r = 0.64) and PCS (r = 0.55) (
< 0.01), as well as MAP
/DBP
for PES (r = 0.6/0.55) and PCS (r = 0.5/0.45) (
< 0.05). A 7% decrease in 1/PTT
predicted a 30% SBP
decrease (0.82, 0.76, 0.76), while a 5.6% increase predicted a 30% SBP
increase (0.75, 0.7, 0.68). A 6.6% decrease in 1/PTT
detected a 30% SBP
decrease (0.81, 0.72, 0.8), while a 4.8% 1/PTT
increase detected a 30% SBP
increase (0.73, 0.64, 0.68).
Non-invasive beat-to-beat PTT via PES/PCS demonstrated significant correlations with IBP and detected significant changes in SBP
. Thus, PES/PCS as a novel sensor technology may augment intraoperative hemodynamic monitoring during major surgery. |
---|---|