A study of noninvasive fractional flow reserve derived from a simplified method based on coronary computed tomography angiography in suspected coronary artery disease
The invasive fractional flow reserve has been considered the gold standard for identifying ischaemia-related stenosis in patients with suspected coronary artery disease. Determining non-invasive FFR based on coronary computed tomographic angiography datasets using computational fluid dynamics tends...
Gespeichert in:
Veröffentlicht in: | Biomedical engineering online 2017-04, Vol.16 (1), p.43-43, Article 43 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The invasive fractional flow reserve has been considered the gold standard for identifying ischaemia-related stenosis in patients with suspected coronary artery disease. Determining non-invasive FFR based on coronary computed tomographic angiography datasets using computational fluid dynamics tends to be a demanding process. Therefore, the diagnostic performance of a simplified method for the calculation of FFR
requires further evaluation.
The aim of this study was to investigate the diagnostic performance of FFR
calculated based on a simplified method by referring to the invasive FFR in patient-specific coronary arteries and clinical decision-making.
Twenty-nine subjects included in this study underwent CCTA before undergoing clinically indicated invasive coronary angiography for suspected coronary artery disease. Pulsatile flow simulation and a novel boundary condition were used to obtain FFR
based on the CCTA datasets. The Pearson correlation, Bland-Altman plots and the diagnostic performance of FFR
and CCTA stenosis were analyzed by comparison to the invasive FFR reference standard. Ischaemia was defined as an FFR or FFR
≤0.80, and anatomically obstructive CAD was defined as a CCTA stenosis >50%.
FFR
and invasive FFR were well correlated (r = 0.742, P = 0.001). Slight systematic underestimation was found in FFR
(mean difference 0.03, standard deviation 0.05, P = 0.001). The area under the receiver-operating characteristic curve was 0.93 for FFR
and 0.75 for CCTA on a per-vessel basis. Per-patient accuracy, sensitivity and specificity were 79.3, 93.7 and 61.5%, respectively, for FFR
and 62.1, 87.5 and 30.7%, respectively, for CCTA. Per-vessel accuracy, sensitivity and specificity were 80.6, 94.1 and 68.4%, respectively, for FFR
and 61.6, 88.2 and 36.8%, respectively, for CCTA.
FFR
derived from pulsatile simulation with a simplified novel boundary condition was in good agreement with invasive FFR and showed better diagnostic performance compared to CCTA, suggesting that the simplified method has the potential to be an alternative and accurate way to assess the haemodynamic characteristics for coronary stenosis. |
---|---|
ISSN: | 1475-925X 1475-925X |
DOI: | 10.1186/s12938-017-0330-2 |