Evaluation of the eazyplex MRSA assay for the rapid detection of Staphylococcus aureus in pleural and synovial fluid

Highlights • The commercial loop-mediated isothermal amplification (LAMP) eazyplex MRSA assay was validated for the detection of Staphylococcus aureus including methicillin-resistant Staphylococcus aureus (MRSA) isolates in pleural and synovial fluid. • This new system appropriately detected a quali...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:International journal of infectious diseases 2017-06, Vol.59 (C), p.65-68
Hauptverfasser: Henares, D, Broton, P, Buyse, X, Latorre, I, de Paz, H.D, Muñoz-Almagro, C
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Highlights • The commercial loop-mediated isothermal amplification (LAMP) eazyplex MRSA assay was validated for the detection of Staphylococcus aureus including methicillin-resistant Staphylococcus aureus (MRSA) isolates in pleural and synovial fluid. • This new system appropriately detected a quality control panel of pleural/synovial fluid samples with DNA of methicillin-susceptible Staphylococcus aureus (MSSA), MRSA, and other pathogens. • The limit of detection was established at 6.4 × 103 CFU/ml for S. aureus and 1.0 × 104 CFU/ml for MRSA. • The diagnostic sensitivity and specificity values were 83.3% and 97.8%, respectively, for the detection of S. aureus , while the mean turnaround time was 70 min. • The eazyplex MRSA assay was found to be a useful test for the rapid detection of S. aureus in invasive samples.