Ultrasensitive Detection of Testosterone Using Microring Resonator with Molecularly Imprinted Polymers
We report ultrasensitive and highly selective detection of testosterone based on microring resonance sensor using molecularly imprinted polymers (MIP). A silicon-on-insulator (SOI) micoring resonator was modified by MIP films (MIPs) on a surface. The MIPs was synthesized by thermopolymerization usin...
Gespeichert in:
Veröffentlicht in: | Sensors (Basel, Switzerland) Switzerland), 2015-12, Vol.15 (12), p.31558-31565 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We report ultrasensitive and highly selective detection of testosterone based on microring resonance sensor using molecularly imprinted polymers (MIP). A silicon-on-insulator (SOI) micoring resonator was modified by MIP films (MIPs) on a surface. The MIPs was synthesized by thermopolymerization using methacrylic acid as functional monomer and ethylene glycol dimethacrylate as crosslinking agent. The concentration of detected testosterone varies from 0.05 ng/mL to 10 ng/mL. The detection limit reaches 48.7 pg/mL. Ultrahigh sensitivity, good specificity and reproducibility have been demonstrated, indicating the great potential of making a cost effective and easy to operate lab-on-Chip and down scaling micro-fluidics devices in biosensing. |
---|---|
ISSN: | 1424-8220 1424-8220 |
DOI: | 10.3390/s151229877 |