An inappropriately-appropriate ICD therapy: What is the mechanism and what interventions can be done?

R-wave double counting is a rare cause of ventricular oversensing that can lead to inappropriate shocks. Optimizing device programming is essential for the avoidance of subsequent inappropriate therapies. R-wave double counting is a rare cause of ventricular oversensing that can lead to inappropriat...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Clinical case reports 2023-10, Vol.11 (10), p.e8082-n/a
Hauptverfasser: Saplaouras, Athanasios, Mililis, Panagiotis, Kariki, Ourania, Bazoukis, George, Batsouli, Athena, Antonakos, Vasileios, Dragasis, Stylianos, Patsiotis, Ilias G, Letsas, Konstantinos P, Efremidis, Michael
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:R-wave double counting is a rare cause of ventricular oversensing that can lead to inappropriate shocks. Optimizing device programming is essential for the avoidance of subsequent inappropriate therapies. R-wave double counting is a rare cause of ventricular oversensing that can lead to inappropriate shocks. We present the case of a female patient, 52-years-old with a history of end-stage hypertrophic cardiomyopathy. The patient suffered an implantable cardioverter-defibrillator (ICD) shock. Cardiac device interrogation revealed ventricular tachycardia (VT) with a cycle length of 420 ms that was misclassified in the ventricular fibrillation (VF) zone owing to R-wave double counting. Optimizing device programming is essential for the avoidance of subsequent inappropriate therapies. Possible therapeutic options are briefly presented in the case.
ISSN:2050-0904
2050-0904
DOI:10.1002/ccr3.8082