Evaluation of dose rate and photon energy dependence of gafchromic EBT3 film irradiating with 6 MV and Co-60 photon beams
Gafchromic films are utilized for two-dimensional dose distribution measurements, especially in radiotherapy. In this study, we investigated a close connection between energy and dose rate of Gafchromic EBT3 films irradiating with 6 MV and Co-60 photon beams over a broad dose range. EBT3 films were...
Gespeichert in:
Veröffentlicht in: | Journal of medical signals and sensors 2019-07, Vol.9 (3), p.204-210 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Gafchromic films are utilized for two-dimensional dose distribution measurements, especially in radiotherapy. In this study, we investigated a close connection between energy and dose rate of Gafchromic EBT3 films irradiating with 6 MV and Co-60 photon beams over a broad dose range. EBT3 films were exposed to 6 MV and Co-60 photon beams using 4 and 2 Gy/min dose rates over a 10-400 cGy dose range. The films were scanned in red, green, and blue channels to obtain the optical density (OD)-dose curves. The OD-dose curves resulted from three-color scans for different photon energies and dose rates were compared by statistical independent t-test. For the radiations of Co-60 and 6 MV photon beams, the highest correlation was obtained between the 2 and 4 Gy/min dose rates with red and green channels, respectively. Moreover, the red channel had a greater OD response per dose value, following the green and blue channels. There was no significant difference between different photon energies' (Co-60 and 6 MV) and dose rates' (2 and 4 Gy/min) dependence on OD-dose response of EBT3 films over a broad domain of radiation dose, except for different photon energies in the blue channel. Our results revealed that the OD-dose response of EBT3 films is independent on photon energies (Co-60 and 6 MV) and dose rate (2 and 4 Gy/min) in the evaluated dose range (10-400 cGy). Therefore, the EBT3 films are suitable, consistent, and reliable instruments for dose measurements in radiotherapy. |
---|---|
ISSN: | 2228-7477 2228-7477 |
DOI: | 10.4103/jmss.JMSS_45_18 |