Volumetric modulated arc therapy based total body irradiation: Workflow and clinical experience with an indexed rotational immobilization system

An indexed rotational immobilization system was developed for supine total body irradiation (TBI). Treatment plans had multi-isocentric volumetric modulated arc therapy (VMAT) beams to the upper body and parallel-opposed fields to the lower body, with a 12 Gy prescription dose to >90% of the body...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Physics and imaging in radiation oncology 2017-10, Vol.4, p.22-25
Hauptverfasser: Ouyang, Luo, Folkerts, Michael, Zhang, You, Hrycushko, Brian, Lamphier, Richard, Lee, Pam, Chambers, Eric, Ramirez, Ezequiel, Reynolds, Robert, Yan, Yulong, Jiang, Steve B., Timmerman, Robert, Desai, Neil, Abdulrahman, Ramzi, Gu, Xuejun
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:An indexed rotational immobilization system was developed for supine total body irradiation (TBI). Treatment plans had multi-isocentric volumetric modulated arc therapy (VMAT) beams to the upper body and parallel-opposed fields to the lower body, with a 12 Gy prescription dose to >90% of the body and mean lung dose ∼8 Gy. In the end-to-end test, point dose measurements had
ISSN:2405-6316
2405-6316
DOI:10.1016/j.phro.2017.11.002