Dry textile electrode for ambulatory monitoring after catheter ablation of atrial fibrillation: A pilot study of simultaneous comparison to the Holter electrocardiogram [version 1; peer review: 1 approved, 1 approved with reservations]

Background: Holter electrocardiogram (ECG) is the gold standard for ambulatory monitoring of atrial fibrillation (AF) but it is insufficient because of its limited recording time. A wearable ECG with a medical-grade dry textile electrode is a promising technology to remedy this limitation. This pilo...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:F1000 research 2022, Vol.11, p.97
Hauptverfasser: Machino, Takeshi, Aonuma, Kazutaka, Komatsu, Yuki, Yamasaki, Hiro, Igarashi, Miyako, Nogami, Akihiko, Ieda, Masaki
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Background: Holter electrocardiogram (ECG) is the gold standard for ambulatory monitoring of atrial fibrillation (AF) but it is insufficient because of its limited recording time. A wearable ECG with a medical-grade dry textile electrode is a promising technology to remedy this limitation. This pilot study aimed to simultaneously compare the wearable and Holter ECGs for ambulatory monitoring in a clinical setting. Methods: This prospective observational study enrolled 18 patients who underwent AF ablation. One day after AF ablation, ambulatory ECG was obtained for three hours simultaneously using both the wearable and Holter ECG devices. Automatic ECG interpretations between devices were compared with correlation and agreement analyses. Results: Simultaneous ECG monitoring demonstrated a comparable analysis time and total heart beats between the two devices. Almost complete correlation and agreement were also demonstrated in all clinically relevant testing aspects except in R-wave amplitude (r = 0.743, p < .001). AF was detected in three patients. AF duration was the same in both ECG devices in two patients with continuous AF. In the remaining patient with intermittent AF, AF duration was shortened by 0.6% with the wearable ECG as compared to that with the Holter ECG. Conclusions: Simultaneous ECG comparison revealed a high consistency between the wearable and Holter ECG devices. The results of this study warrant further clinical studies for long-term monitoring of ambulatory ECG after AF ablation.
ISSN:2046-1402
2046-1402
DOI:10.12688/f1000research.75712.1