In vitro assessment of internal implant-abutment connections with different cone angles under static loading using synchrotron-based radiation

The stability of implant-abutment connection is crucial to minimize mechanical and biological complications. Therefore, an assessment of the microgap behavior and abutment displacement in different implant-abutment designs was performed. Four implant systems were tested, three with a conical implant...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:BMC oral health 2024-03, Vol.24 (1), p.396-396, Article 396
Hauptverfasser: Angermair, Johannes, Iglhaut, Gerhard, Meyenberg, Konrad, Wiest, Wolfram, Rack, Alexander, Zabler, Simon, Fretwurst, Tobias, Nelson, Katja, Kernen, Florian
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The stability of implant-abutment connection is crucial to minimize mechanical and biological complications. Therefore, an assessment of the microgap behavior and abutment displacement in different implant-abutment designs was performed. Four implant systems were tested, three with a conical implant-abutment connection based on friction fit and a cone angle 
ISSN:1472-6831
1472-6831
DOI:10.1186/s12903-024-04156-2