Comparison of Two Commercial Cyclosporin Assays

A new Dade Behring Cyclosporin A (CsA) assay has been made commercially accessible recently. The aim of this study was to examine the analytical performance of this assay and the correlation with the Abbott Diagnostics assay. The accuracy and precision within-run were good, but precision between-run...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of medical biochemistry 2007-01, Vol.26 (1), p.33-37
Hauptverfasser: Vuceljic, Marina, Lalic, Natasa, Mijuskovic, Zoran, Pejovic, Janko
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A new Dade Behring Cyclosporin A (CsA) assay has been made commercially accessible recently. The aim of this study was to examine the analytical performance of this assay and the correlation with the Abbott Diagnostics assay. The accuracy and precision within-run were good, but precision between-run slightly exceeded the manufacturer's recommendation on one quality-control material. Dade Behring CsA whole blood results from the kidney recipients (n = 104) were lower (p
ISSN:1452-8258
1452-8266
DOI:10.2478/v10011-007-0006-8