Evaluation of Aerosol Electrospray Analysis of Metal-on-Metal Wear Particles from Simulated Total Joint Replacement

Wear is a common cause for aseptic loosening in artificial joints. The purpose of this study was to develop an automated diagnostical method for identification of the number and size distribution of wear debris. For this purpose, metal debris samples were extracted from a hip simulator and then anal...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Sensors (Basel, Switzerland) Switzerland), 2019-08, Vol.19 (17), p.3751
Hauptverfasser: von Skrbensky, Gobert, Mühlbacher, Karoline, Benca, Emir, Kolb, Alexander, Windhager, Reinhard, Reischl, Georg, Reinisch, Georg
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Wear is a common cause for aseptic loosening in artificial joints. The purpose of this study was to develop an automated diagnostical method for identification of the number and size distribution of wear debris. For this purpose, metal debris samples were extracted from a hip simulator and then analyzed by the electrospray method combined with a differential mobility analyzer, allowing particle detection ranging from several nanometers up to 1 µm. Wear particles were identified with a characteristic peak at 15 nm. The electrospray setup was successfully used and validated for the first time to characterize wear debris from simulated total joint replacement. The advantages of this diagnostic method are its time- and financial efficiency and its suitability for testing of different materials.
ISSN:1424-8220
1424-8220
DOI:10.3390/s19173751