A Novel Solution for the High Defibrillation Threshold in Patients with a DF-4 Lead: Adding a High Voltage Adaptor/Splitter

A high defibrillation threshold occurs in approximately 6% of implants. The defibrillation threshold can be improved by addition of a defibrillation lead. However, the DF-4 high energy ICD header precludes the addition of a defibrillation lead. Here we report on use of a new high voltage adaptor/spl...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Indian pacing and electrophysiology journal 2014-05, Vol.14 (3), p.152-156
Hauptverfasser: Inbar, Shmuel, Seethala, Srikanth
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A high defibrillation threshold occurs in approximately 6% of implants. The defibrillation threshold can be improved by addition of a defibrillation lead. However, the DF-4 high energy ICD header precludes the addition of a defibrillation lead. Here we report on use of a new high voltage adaptor/splitter that enables the addition of an extra defibrillation lead.