Comparing the novel microstream and the traditional mainstream method of end-tidal CO2 monitoring with respect to PaCO2 as gold standard in intubated critically ill children
The objective of this study was to evaluate a novel microstream method by comparison with PaCO 2 and the more standard mainstream capnometer in intubated pediatric patients. We hypothesized that the novel microstream method would superior compared to the traditional mainstream method in predicting P...
Gespeichert in:
Veröffentlicht in: | Scientific reports 2020-12, Vol.10 (1), p.22042-22042, Article 22042 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The objective of this study was to evaluate a novel microstream method by comparison with PaCO
2
and the more standard mainstream capnometer in intubated pediatric patients. We hypothesized that the novel microstream method would superior compared to the traditional mainstream method in predicting PaCO
2
. This was a prospective single-center comparative study. The study was carried out on 174 subjects with a total of 1338 values for each method. Data were collected prospectively from mainstream and microstream capnometer simultaneously and compared with PaCO
2
results. Although both mainstream PetCO
2
(mainPetCO
2
) and microstream PetCO
2
(microPetCO
2
) were moderately correlated (r = 0.63 and r = 0.68, respectively) with PaCO
2
values, mainPetCO
2
was in better agreement with PaCO
2
in all subjects (bias ± precision values of 3.8 ± 8.9 and 7.3 ± 8.2 mmHg, respectively). In those with severe pulmonary disease, the mainPetCO
2
and microPetCO
2
methods were highly correlated with PaCO
2
(r = 0.80 and r = 0.81, respectively); however, the biases of both methods increased (14.8 ± 9.1 mmHg and 16.2 ± 9.0 mmHg, respectively). In cases with increased physiologic dead space ventilation, the agreement levels of mainPetCO
2
and microPetCO
2
methods became distorted (bias ± precision values of 20.9 ± 11.2 and 25.0 ± 11.8 mm Hg, respectively) even though mainPetCO
2
and microPetCO
2
were highly correlated (r = 0.78 and r = 0.78, respectively). It was found that the novel microstream capnometer method for PetCO
2
measurements provided no superiority to the traditional mainstream method. Both capnometer methods may be useful in predicting the trend of PaCO
2
due to significant correlations with the gold standard measurement in cases with severe pulmonary disease or increased physiological dead space –despite reduced accuracy. |
---|---|