Clinical Performance of Three Commercial SARS‐CoV‐2 Rapid Antigen Tests for Community-Dwelling Individuals in a Tropical Setting
During the second year of the COVID-19 pandemic, the use of Rapid Diagnosis Antigen Tests (RDAgTs) for SARS-CoV-2 detection has substantially increased as some of the brands available in the market were certified for clinical use by international regulatory agencies. RDAgTs are a fast and cheap tool...
Gespeichert in:
Veröffentlicht in: | Frontiers in cellular and infection microbiology 2022-07, Vol.12, p.832235-832235 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | During the second year of the COVID-19 pandemic, the use of Rapid Diagnosis Antigen Tests (RDAgTs) for SARS-CoV-2 detection has substantially increased as some of the brands available in the market were certified for clinical use by international regulatory agencies. RDAgTs are a fast and cheap tool for SARS-CoV-2 surveillance with great potential to improve testing capacities in middle- and low-income countries compared to the gold standard RT-qPCR. However, as the clinical performance of RDAgTs has been shown to vary greatly between the commercial brands available, evaluation studies are necessary. Moreover, the available evaluation has been done in high-income countries while SARS-CoV-2 transmission is also actively happening in developing countries, many of which are located in tropical latitudes where cross-reactivity with other infectious agents is highly prevalent, which could compromise RDAgT specificity. Moreover, unreported mutations and/or new SARS-CoV-2 variants may compromise RDAgT sensitivity as genomic surveillance is limited in these settings. Here we describe a multicenter and manufacturer‐independent evaluation of the clinical performance and analytical sensitivity of three different RDAgTs brands available in South America from three companies, Rapigen (South Korea), SD-Biosensor (South Korea), and Certest (Spain), compared to the gold standard RT-qPCR. A total number of 1,646 nasopharyngeal swabs from community-dwelling individuals were included in the study, and 379 of them were SARS-CoV-2 positive by RT-qPCR. The overall sensitivity for each RDAgT was 79% (IC95%: 72 - 86.2), 64.2% (IC95%: 56.7 - 71.6), and 45.8% (IC95%: 35.8 - 55.8) for SD-Biosensor, Certest, and Rapigen, respectively. The overall specificity for each RDAgT was 100%, 97.7% (IC95%: 96.8 - 98.6), and 100% for SD-Biosensor, Certest, and Rapigen, respectively. However, the limit of detection (LoD) to achieve a sensitivity over 90% was substantially lower for Certest RDAgT (10
2
copies/uL) compared to SD-Biosensor (10
3
copies/uL) or Rapigen (10
6
copies/uL) RDAgTs, considering that the gold standard RT-qPCR method used in this study has a high sensitivity of 97.7% and low LoD of 5 copies/uL. Additionally, the Certest RDAgT also showed an improved sensitivity up to 79.7% (IC95%: 70.2 – 89.2) for symptomatic individuals. Finally, the slight reduction in specificity for Certest RDAgTs was only associated with one of the laboratories performing this study, pointing out the ne |
---|---|
ISSN: | 2235-2988 2235-2988 |
DOI: | 10.3389/fcimb.2022.832235 |