Investigation of the effect of load rate on creep and hysteresis errors in strain gauge and piezoelectric force transducer

This study includes measurements carried out according to ISO 376 international standard and DKD-R 3-9 guideline by using different low pass filters and continuous data acquisition, with the purpose of comprehensively investigate creep, influence of load rate and influence of load rate on hysteresis...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Measurement. Sensors 2022-08, Vol.22, p.100374, Article 100374
Hauptverfasser: Dizdar, Haldun, Aydemir, Bulent, Vatan, Cemal
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:This study includes measurements carried out according to ISO 376 international standard and DKD-R 3-9 guideline by using different low pass filters and continuous data acquisition, with the purpose of comprehensively investigate creep, influence of load rate and influence of load rate on hysteresis in force sensors that have two different measuring principles. It has been revealed and evaluated the contribution of these parameters to the measurement uncertainty.
ISSN:2665-9174
2665-9174
DOI:10.1016/j.measen.2022.100374