Application of error classification model using indices based on dose distribution for characteristics evaluation of multileaf collimator position errors
This study aims to evaluate the specific characteristics of various multileaf collimator (MLC) position errors that are correlated with the indices using dose distribution . The dose distribution was investigated using the gamma, structural similarity, and dosiomics indices. Cases from the American...
Gespeichert in:
Veröffentlicht in: | Scientific reports 2023-07, Vol.13 (1), p.11027-11027, Article 11027 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | This study aims to evaluate the specific characteristics of various multileaf collimator (MLC) position errors that are correlated with the indices using dose distribution
.
The dose distribution was investigated using the gamma, structural similarity, and dosiomics indices. Cases from the American Association of Physicists in Medicine Task Group 119 were planned, and systematic and random MLC position errors were simulated. The indices were obtained from distribution maps and statistically significant indices were selected. The final model was determined when all values of the area under the curve, accuracy, precision, sensitivity, and specificity were higher than 0.8 (
p
0.9. Furthermore, the results of the DVH were related to dosiomics analysis in that it reflects the characteristics of the MLC position error. It was also shown that dosiomics analysis could provide important information on localized dose-distribution differences in addition to DVH information. |
---|---|
ISSN: | 2045-2322 2045-2322 |
DOI: | 10.1038/s41598-023-35570-1 |