Riata lead failure presenting as a life-threatening electrical storm: A novel manifestation of electrical lead failure

Abstract The Riata and Riata ST family of implantable cardioverter-defibrillator leads are prone to a unique type of structural failure involving exteriorization of the conductor cables, which may present as electrical failure. We report a mode of lead failure that occurred in a patient with a Riata...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of arrhythmia 2014-12, Vol.30 (6), p.525-528
Hauptverfasser: Shenthar, Jayaprakash, MD, DM, FACC, Ghanta, Somasekhar, MD, DM, Rai, Maneesh K., MD, DM, Walia, Rohit, MD, DM
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Abstract The Riata and Riata ST family of implantable cardioverter-defibrillator leads are prone to a unique type of structural failure involving exteriorization of the conductor cables, which may present as electrical failure. We report a mode of lead failure that occurred in a patient with a Riata 1570 series dual coil 8F lead. In this case, the first appropriate shock for ventricular fibrillation resulted in noise, that in turn led to recurrent inappropriate shocks and proarrhythmia, that clinically mimicked a life-threatening electrical storm.