Study on type-testing of a manual TLD-reader for dosimetry programs

In a radiation individual monitoring program, the type testing of measuring devices is a great important part of the quality management system. The IEC-62387 standard applies to dosimetry systems that measure external photon and/or beta radiation within limited ranges of the associated physical para...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of nuclear research and applications 2022-01, Vol.2 (1), p.64-70
Hauptverfasser: Hosseini Pooya, S.M., Rezaeian, P., Edalatkhah, E.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:In a radiation individual monitoring program, the type testing of measuring devices is a great important part of the quality management system. The IEC-62387 standard applies to dosimetry systems that measure external photon and/or beta radiation within limited ranges of the associated physical parameters. In this work, a type-testing program was conducted for a manual thermoluminescence dosimetry (TLD) reader employing the IEC-62387  radiation and environmental performance criteria. The uncertainty of non-linearity of the response of the dosimetry system in a range of 0.7-850 mSv was obtained between -15% and + 17%, which fulfilled the IEC standard range of -16% to +18%. Furthermore, the total uncertainty of all reader tests was measured to be 12%, which was less than the criteria of 20% in the IEC standard. Thus, it can be concluded that the TLD reader met all requirements of the IEC standard for the reader-tests by an appropriate margin.
ISSN:2783-3402
2783-3402
DOI:10.24200/jon.2022.1013