Impact of image averaging on vessel detection using optical coherence tomography angiography in eyes with macular oedema and in healthy eyes
PurposeTo assess the repeatability of multiple automatic vessel density (VD) measurements and the effect of image averaging on vessel detection by optical coherence tomography angiography (OCTA).MethodsAn observational study was conducted in a series of healthy volunteers and patients with macular o...
Gespeichert in:
Veröffentlicht in: | PloS one 2021-01, Vol.16 (10), p.e0257859 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | PurposeTo assess the repeatability of multiple automatic vessel density (VD) measurements and the effect of image averaging on vessel detection by optical coherence tomography angiography (OCTA).MethodsAn observational study was conducted in a series of healthy volunteers and patients with macular oedema. Five sequential OCTA images were acquired for each eye using the OptoVue HD device. The effect of the averaging of the 5 acquisitions on vessel detection was analysed quantitatively using a pixel-by-pixel automated analysis. In addition, two independent retina experts qualitatively assessed the change in vessel detection in averaged images segmented in 9 boxes and compared to the first non-averaged image.ResultsThe automatic VD measurement in OCTA images showed a good repeatability with an overall mean intra-class correlation coefficient (ICC) of 0.924. The mean ICC was higher in healthy eyes compared to eyes with macular oedema (0.877 versus 0.960; p < 0.001) and in the superficial vascular plexus versus the deep vascular complex (0.967 versus 0.888; p = 0.001). The quantitative analysis of the effect of the averaging showed that averaged images had a mean gain of 790.4 pixels/box, located around or completing interruptions in the vessel walls, and a mean loss of 727.2 pixels/box. The qualitative analysis of the averaged images showed that 99.6% of boxes in the averaged images had a gain in vessel detection (i.e., vessels detected in the averaged image but not in the non-averaged image). The loss of pixels was due to a reduction in background noise and motion artifacts in all cases and no case of loss of vessel detection was observed.ConclusionThe automatic VD measurement using the OptoVue HD device showed a good repeatability in 5 acquisitions in a row setting. Averaging images increased vessel detection, and in about a third of boxes, decreased the background noise, both in healthy eyes and, in a greater proportion, in eyes with macular oedema. |
---|---|
ISSN: | 1932-6203 |
DOI: | 10.1371/journal.pone.0257859 |