A comprehensive quality assurance program for four-dimensional computed tomography in radiotherapy

This study aimed to develop and validate a comprehensive, reproducible and automatic 4DCT Quality Assurance(QA) workflow (QAMotion) that evaluates image accuracy across various regular and irregular breathing patterns. Volume and amplitude deviations, CT number accuracy, and spatial integrity were u...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Physics and imaging in radiation oncology 2023-07, Vol.27, p.100475-100475, Article 100475
Hauptverfasser: Bakkali Tahiri, Jinane, Kyndt, Martin, Dhont, Jennifer, Gulyban, Akos, Szkitsak, Juliane, Bogaert, Evelien, Reynaert, Nick, Burghelea, Manuela
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:This study aimed to develop and validate a comprehensive, reproducible and automatic 4DCT Quality Assurance(QA) workflow (QAMotion) that evaluates image accuracy across various regular and irregular breathing patterns. Volume and amplitude deviations, CT number accuracy, and spatial integrity were used as evaluation metrics. For repeatability tests, tolerances were respected with a mean CT number deviation