In vivo axial load-share ratio measurement using a novel hexapod system for safe external fixator removal

External fixation is widely used in the treatment of traumatic fractures; however, orthopedic surgeons encounter challenges in deciding the optimal time for fixator removal. The axial load-share ratio (LS) of the fixator is a quantitative index to evaluate the stiffness of callus healing. This paper...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:BMC musculoskeletal disorders 2024-05, Vol.25 (1), p.353-353, Article 353
Hauptverfasser: Liu, Sida, Lu, Lin, Chen, Tao, Liu, Yanshi, Wei, Dong, Miao, Jun, Yu, Defu, Fu, Xuefei
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:External fixation is widely used in the treatment of traumatic fractures; however, orthopedic surgeons encounter challenges in deciding the optimal time for fixator removal. The axial load-share ratio (LS) of the fixator is a quantitative index to evaluate the stiffness of callus healing. This paper introduces an innovative method for measuring the LS and assesses the method's feasibility and efficacy. Based on a novel hexapod LS-measurement system, the proposed method is to improve the convenience and precision of measuring LS in vivo, hence facilitating the safe removal of external fixators. A novel hexapod system is introduced, including its composition, theoretical model, and method for LS measurement. We conducted a retrospective study on 82 patients with tibial fractures treated by the Taylor Spatial Frame in our hospital from September 2018 to June 2020, of which 35 took LS measurements with our novel method (Group I), and 47 were with the traditional method (Group II). The external fixator was removed when the measurement outcome (LS 
ISSN:1471-2474
1471-2474
DOI:10.1186/s12891-024-07440-y