Stability indicating RP-HPLC method for simultaneous determination of gatifloxacin and dexamethasone in binary combination

In this study, conditions were optimized for development of a simple RP-HPLC method for simultaneous analysis of gatifloxacin and dexamethasone in different matrices like pharmaceuticals, human serum and urine. Good separation of gatifloxacin and dexamethasone from the induced degradation products w...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Brazilian Journal of Pharmaceutical Sciences 2017-01, Vol.53 (1)
Hauptverfasser: Razzaq, Syed Naeem, Ashfaq, Muhammad, Khan, Islam Ullah, Mariam, Irfana, Razzaq, Syed Saleem, Mustafa, Ghulam, Zubair, Muhammad
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:In this study, conditions were optimized for development of a simple RP-HPLC method for simultaneous analysis of gatifloxacin and dexamethasone in different matrices like pharmaceuticals, human serum and urine. Good separation of gatifloxacin and dexamethasone from the induced degradation products was accomplished using C8 as stationary phase; 0.02 M phosphate buffer (pH 3.0) and methanol (42:58 v/v) as mobile phase. The concentration was measured with DAD at 270 nm. Linearity was observed in the range of 0.000040-0.000280 mol/L for gatifloxacin (r2≥0.999) and 0.000013-0.000091 mol/L for dexamethasone (r2≥0.999). Both the analyte peaks were completely separated from the peaks of induced degradation products as indicated by the peak purity index (≥0.9999 for both analytes). The optimized method is recommended to be used for concurrent analysis of gatifloxacin and dexamethasone in different matrices.
ISSN:1984-8250
2175-9790
1984-8250
2175-9790
DOI:10.1590/s2175-97902017000115177