Evaluation of the Diagnostic Accuracy of Nasal Cavity and Nasopharyngeal Swab Specimens for SARS-CoV-2 Detection via Rapid Antigen Test According to Specimen Collection Timing and Viral Load

The rapid diagnosis of SARS-CoV-2 is an essential aspect in the detection and control of the spread of COVID-19. We evaluated the accuracy of the rapid antigen test (RAT) using samples from the nasal cavity and nasopharynx based on sample collection timing and viral load. We enrolled 175 patients, o...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Diagnostics (Basel) 2022-03, Vol.12 (3), p.710
Hauptverfasser: Lee, Seungjun, Widyasari, Kristin, Yang, Hye-Ryun, Jang, Jieun, Kang, Taejoon, Kim, Sunjoo
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The rapid diagnosis of SARS-CoV-2 is an essential aspect in the detection and control of the spread of COVID-19. We evaluated the accuracy of the rapid antigen test (RAT) using samples from the nasal cavity and nasopharynx based on sample collection timing and viral load. We enrolled 175 patients, of which 71 patients and 104 patients had tested positive and negative, respectively, based on real time-PCR. Nasal cavity and nasopharyngeal swab samples were tested using STANDARD Q COVID-19 Ag tests (Q Ag, SD Biosensor, Korea). The sensitivity of the Q Ag test was 77.5% (95% confidence interval [CI], 67.8−87.2%) for the nasal cavity and 81.7% (95% [CI, 72.7−90.7%) for the nasopharyngeal specimens. The RAT results showed a substantial agreement between the nasal cavity and nasopharyngeal specimens (Cohen’s kappa index = 0.78). The sensitivity of the RAT for nasal cavity specimens exceeded 89% for
ISSN:2075-4418
2075-4418
DOI:10.3390/diagnostics12030710