Feasibility of Atrial AutoCapture™ to Detect Atrial Evoked Response: Experience from 102 Patients Implanted with Dual-chamber Pacemakers
Atrial AutoCapture™ (ACap™) was a new technological development that confirmed atrial capture by analyzing evoked response (ER) with a new method - paced depolarization integral ER detection - and optimized energy output to changes in the stimulation threshold. The purpose of this study was to evalu...
Gespeichert in:
Veröffentlicht in: | Chinese medical journal 2017-06, Vol.130 (12), p.1411-1417 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Atrial AutoCapture™ (ACap™) was a new technological development that confirmed atrial capture by analyzing evoked response (ER) with a new method - paced depolarization integral ER detection - and optimized energy output to changes in the stimulation threshold. The purpose of this study was to evaluate the clinical performance of ACap™ function.
This was a prospective, observational, nonrandomized two-center study. Between November 2008 and August 2014, 102 patients were enrolled from two different institutions. Data were collected by case report forms at enrollment, hospital discharge, and in-office follow-ups scheduled at 1, 2, 3, 6, and 12 months postimplantation.
Ambulatory ACap™ function started to become available for 20.6% of patients at 1 day, then progressed to 30.4% at 7 days, 38.6% at 1 month, 41.6% at 2 months, 47.5% at 3 months, 53.5% at 6 months, and 63.4% at 1 year. The cause of the unsuccessful attempts to perform ACap™ threshold was ER/polarization |
---|---|