A measuring method of standard resistance and analysis of measuring results

DC standard resistance is widely used in precision measurement, verification, traceability and other fields due to its high accuracy and good stability. First-class resistance standard has used to measure it based on the bridge turn equilibrium principle. According to JJF1059.1–2012 Evaluation and e...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Measurement. Sensors 2021-12, Vol.18, p.100207, Article 100207
Hauptverfasser: Guan, Zexin, XinyanWang, Mei Yang, Li, Wenqiang, Sun, Xiaoyan, Wang, Dalong, Wang, Shang
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:DC standard resistance is widely used in precision measurement, verification, traceability and other fields due to its high accuracy and good stability. First-class resistance standard has used to measure it based on the bridge turn equilibrium principle. According to JJF1059.1–2012 Evaluation and expression of uncertainty in measurement, the uncertainty of measurement results has been evaluated on uncertainty of the transfer standard, the basic error, sensitivity, load power, tank temperature, thermoelectric power and staff. This paper analyzes the uncertainty of the measurement results which will useful to measurement uncertainty assessment of standard resistance in future testing work.
ISSN:2665-9174
2665-9174
DOI:10.1016/j.measen.2021.100207