Preoperative measurement vs intraoperative aberrometry for the selection of intraocular lens sphere power in normal eyes
This study aimed to assess the value of intraoperative aberrometry (IA) in determining the intraocular lens (IOL) sphere power in eyes with no previous ocular surgery. We conducted a retrospective review of patients who underwent uncomplicated cataract surgery where standard preoperative (Preop) mea...
Gespeichert in:
Veröffentlicht in: | Clinical ophthalmology (Auckland, N.Z.) N.Z.), 2017-01, Vol.11, p.923-929 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | This study aimed to assess the value of intraoperative aberrometry (IA) in determining the intraocular lens (IOL) sphere power in eyes with no previous ocular surgery.
We conducted a retrospective review of patients who underwent uncomplicated cataract surgery where standard preoperative (Preop) measurements and IA were performed. Calculated IOL sphere powers and postoperative refractions, both actual and theoretical, were compared based on the measurement method and lens type; lens types included multifocal, toric and aspheric single-vision non-toric IOLs.
A total of 160 eyes of 112 patients were analyzed. The Preop lens power calculated was the same as the IA lens power 46% of the time, though this percentage was lower for multifocal IOLs. Across all lens types, there was a statistically significant bias (chi-square test, |
---|---|
ISSN: | 1177-5467 1177-5483 1177-5483 |
DOI: | 10.2147/OPTH.S135659 |