Performance and safety of the PRICO closed-loop oxygen saturation targeting system in neonates: pragmatic multicentre cross-over study (TarOx Study)
ObjectiveThis study aims to evaluate the performance of the fabian-Predictive-Intelligent-Control-of-Oxygenation (PRICO) system for automated control of the fraction of inspired oxygen (FiO2).DesignMulticentre randomised cross-over study.SettingFive neonatal intensive care units experienced with aut...
Gespeichert in:
Veröffentlicht in: | BMJ paediatrics open 2024-07, Vol.8 (1), p.e002583 |
---|---|
Hauptverfasser: | , , , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | ObjectiveThis study aims to evaluate the performance of the fabian-Predictive-Intelligent-Control-of-Oxygenation (PRICO) system for automated control of the fraction of inspired oxygen (FiO2).DesignMulticentre randomised cross-over study.SettingFive neonatal intensive care units experienced with automated control of FiO2 and the fabian ventilator.Patients39 infants: median gestational age of 27 weeks (IQR: 26–30), postnatal age 7 days (IQR: 2–17), weight 1120 g (IQR: 915–1588), FiO2 0.32 (IQR: 0.22–0.43) receiving both non-invasive (27) and invasive (12) respiratory support.InterventionRandomised sequential 24-hour periods of automated and manual FiO2 control.Main outcome measuresProportion (%) of time in normoxaemia (90%–95% with FiO2>0.21 and 90%–100% when FiO2=0.21) was the primary endpoint. Secondary endpoints were severe hypoxaemia (98% with FiO2>0.21) and prevalence of episodes ≥60 s at these two SpO2 extremes.ResultsDuring automated control, subjects spent more time in normoxaemia (74%±22% vs 51%±22%, p |
---|---|
ISSN: | 2399-9772 2399-9772 |