Avatar-based versus conventional vital sign display in a central monitor for monitoring multiple patients: a multicenter computer-based laboratory study
Maintaining adequate situation awareness is crucial for patient safety. Previous studies found that the use of avatar-based monitoring (Visual Patient Technology) improved the perception of vital signs compared to conventional monitoring showing numerical and waveform data; and was further associate...
Gespeichert in:
Veröffentlicht in: | BMC medical informatics and decision making 2020-02, Vol.20 (1), p.26-26, Article 26 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Maintaining adequate situation awareness is crucial for patient safety. Previous studies found that the use of avatar-based monitoring (Visual Patient Technology) improved the perception of vital signs compared to conventional monitoring showing numerical and waveform data; and was further associated with a reduction of perceived workload. In this study, we aimed to evaluate the effectiveness of Visual Patient Technology on perceptive performance and perceived workload when monitoring multiple patients at the same time, such as in central station monitors in intensive care units or operating rooms.
A prospective, within-subject, computer-based laboratory study was performed in two tertiary care hospitals in Switzerland in 2018. Thirty-eight physician and nurse anesthetists volunteered for the study. The participants were shown four different central monitor scenarios in sequence, where each scenario displayed two critical and four healthy patients simultaneously for 10 or 30 s. After each scenario, participants had to recall the vital signs of the critical patients. Perceived workload was assessed with the National Aeronautics and Space Administration Task-Load-Index (NASA TLX) questionnaire.
In the 10-s scenarios, the median number of remembered vital signs significantly improved from 7 to 11 using avatar-based versus conventional monitoring with a mean of differences of 4 vital signs, 95% confidence interval (CI) 2 to 6, p |
---|---|
ISSN: | 1472-6947 1472-6947 |
DOI: | 10.1186/s12911-020-1032-4 |