Impact of Dual-Energy Computed Tomography (DECT) Postprocessing Protocols on Detection of Monosodium Urate (MSU) Deposits in Foot Tendons of Cadavers
To evaluate two different dual-energy computed tomography (DECT) post-processing protocols for the detection of MSU deposits in foot tendons of cadavers with verification by polarizing light microscopy as the gold standard. A total of 40 embalmed cadavers (15 male; 25 female; median age, 82 years; m...
Gespeichert in:
Veröffentlicht in: | Diagnostics (Basel) 2023-06, Vol.13 (13), p.2208 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | To evaluate two different dual-energy computed tomography (DECT) post-processing protocols for the detection of MSU deposits in foot tendons of cadavers with verification by polarizing light microscopy as the gold standard.
A total of 40 embalmed cadavers (15 male; 25 female; median age, 82 years; mean, 80 years; range, 52-99; SD ± 10.9) underwent DECT to assess MSU deposits in foot tendons. Two postprocessing DECT protocols with different Hounsfield unit (HU) thresholds, 150/500 (=established) versus 120/500 (=modified). HU were applied to dual source acquisition with 80 kV for tube A and 140 kV for tube B. Six fresh cadavers (4 male; 2 female; median age, 78; mean, 78.5; range 61-95) were examined by DECT. Tendon dissection of 2/6 fresh cadavers with positive DECT 120 and negative DECT 150 studies were used to verify MSU deposits by polarizing light microscopy.
The tibialis anterior tendon was found positive in 57.5%/100% (DECT 150/120), the peroneus tendon in 35%/100%, the achilles tendon in 25%/90%, the flexor halluces longus tendon in 10%/100%, and the tibialis posterior tendon in 12.5%/97.5%. DECT 120 resulted in increased tendon MSU deposit detection, when DECT 150 was negative, with an overall agreement between DECT 150 and DECT 120 of 80% (
= 0.013). Polarizing light microscope confirmed MSU deposits detected only by DECT 120 in the tibialis anterior, the achilles, the flexor halluces longus, and the peroneal tendons.
The DECT 120 protocol showed a higher sensitivity when compared to DECT 150. |
---|---|