Cobalt toxicity after revision to a metal-on-polyethylene total hip arthroplasty for fracture of ceramic acetabular component
Abstract Cobalt metallosis after revision metal-on-polyethylene total hip arthroplasty for catastrophic failure of ceramic components is uncommon but a potentially devastating complication. Common findings associated with heavy metal toxicity include cardiomyopathy, hypothyroidism, skin rashes, visu...
Gespeichert in:
Veröffentlicht in: | Arthroplasty today 2015-12, Vol.1 (4), p.89-91 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Abstract Cobalt metallosis after revision metal-on-polyethylene total hip arthroplasty for catastrophic failure of ceramic components is uncommon but a potentially devastating complication. Common findings associated with heavy metal toxicity include cardiomyopathy, hypothyroidism, skin rashes, visual disturbances, hearing changes, polycythemia, weakness, fatigue, cognitive deterioration, and neuropathy. We report a case of a 57-year-old woman who presented with complaints of progressively worsening hip pain, fatigue, memory loss, lower extremity sensory loss, persistent tachycardia, and ocular changes 5 years after synovectomy and revision of a failed ceramic-on-ceramic total hip arthroplasty to metal-on-polyethylene components. A cobalt level of 788.1 ppb and chromium level of 140 ppb were found on presentation and subsequently decreased to 468.8 ppb and 105.9 ppb, respectively, 2 weeks after revision to a ceramic-on-polyethylene total hip arthroplasty. Improvement of symptoms accompanied this decrease in cobalt and chromium levels. Revision of failed ceramic arthroplasties with later-generation ceramics to avoid this potential complication is recommended. |
---|---|
ISSN: | 2352-3441 2352-3441 |
DOI: | 10.1016/j.artd.2015.09.002 |