Analysis of Stress and Fracture Strength of Zirconia Implants after Cyclic Loading
The aim of this study was to manufacture a new zirconia implant (Y-TZP) and to evaluate the fracture strength and stresses generated after cyclic loading compared to titanium implants. Thirty-three zirconia and titanium implants were fabricated. The stress generated around the implants was observed...
Gespeichert in:
Veröffentlicht in: | Materials research (São Carlos, São Paulo, Brazil) São Paulo, Brazil), 2015-10, Vol.18 (5), p.1082-1088 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The aim of this study was to manufacture a new zirconia implant (Y-TZP) and to evaluate the fracture strength and stresses generated after cyclic loading compared to titanium implants. Thirty-three zirconia and titanium implants were fabricated. The stress generated around the implants was observed by photoelastic analysis. The implants were then separated into groups (n = 10): Control (no treatment), mechanical fatigue and thermomechanical fatigue. All groups were submitted to a single cantilever flexure test to measure the material fracture strength. The results (2-way ANOVA, Bonferroni test, p |
---|---|