Real‐Time Ventricular Fibrillation Amplitude‐Spectral Area Analysis to Guide Timing of Shock Delivery Improves Defibrillation Efficacy During Cardiopulmonary Resuscitation in Swine
Background The ventricular fibrillation amplitude spectral area (AMSA) predicts whether an electrical shock could terminate ventricular fibrillation and prompt return of spontaneous circulation. We hypothesized that AMSA can guide more precise timing for effective shock delivery during cardiopulmona...
Gespeichert in:
Veröffentlicht in: | Journal of the American Heart Association 2017-11, Vol.6 (11), p.n/a |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Background
The ventricular fibrillation amplitude spectral area (AMSA) predicts whether an electrical shock could terminate ventricular fibrillation and prompt return of spontaneous circulation. We hypothesized that AMSA can guide more precise timing for effective shock delivery during cardiopulmonary resuscitation.
Methods and Results
Three shock delivery protocols were compared in 12 pigs each after electrically induced ventricular fibrillation, with the duration of untreated ventricular fibrillation evenly stratified into 6, 9, and 12 minutes: AMSA‐Driven (AD), guided by an AMSA algorithm; Guidelines‐Driven (GD), according to cardiopulmonary resuscitation guidelines; and Guidelines‐Driven/AMSA‐Enabled (GDAE), as per GD but allowing earlier shocks upon exceeding an AMSA threshold. Shocks delivered using the AD, GD, and GDAE protocols were 21, 40, and 62, with GDAE delivering only 2 AMSA‐enabled shocks. The corresponding 240‐minute survival was 8/12, 6/12, and 2/12 (log‐rank test, P=0.035) with AD exceeding GDAE (P=0.026). The time to first shock (seconds) was (median [Q1–Q3]) 272 (161–356), 124 (124–125), and 125 (124–125) (P |
---|---|
ISSN: | 2047-9980 2047-9980 |
DOI: | 10.1161/JAHA.117.006749 |