A Design of Short Dead-Time RF Coil and RF Switch for Low-Field NMR
Based on the ring-down signal generation principle of nuclear magnetic resonance (NMR), we presented here a design scheme of transmitting and receiving separated short dead-time radio frequency (RF) coil composed of loop-gap resonator (LGR) coil and solenoid coil in low-field NMR system, to improve...
Gespeichert in:
Veröffentlicht in: | Bopuxue zazhi 2021-03, Vol.38 (1), p.1-11 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | chi |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Based on the ring-down signal generation principle of nuclear magnetic resonance (NMR), we presented here a design scheme of transmitting and receiving separated short dead-time radio frequency (RF) coil composed of loop-gap resonator (LGR) coil and solenoid coil in low-field NMR system, to improve transmission efficiency by optimizing tuning and matching network. We also designed a fast RF switch and driver to fulfill the requirement of the RF coil. Based on the simulation results, the short dead-time RF coil was fabricated and applied to a self-developed 9.51 MHz portable low-field NMR spectrometer. The NMR experiments showed that the dead-time was reduced to less than 10 μs, which verified the feasibility of the design scheme. |
---|---|
ISSN: | 1000-4556 1000-4556 |
DOI: | 10.11938/cjmr20202823 |