Comparing urine point-of-care tests to screen preeclampsia: Congo-red dot paper test versus dipstick urinalysis

To compare the urine Congo-red dot paper test (CRD) with dipstick urinalysis to screen preeclampsia (PE). A total of 409 paired spot urine samples were obtained prospectively from women with suspected pre-eclampsia attending for routine hospital visits. Congo-red dot paper test and dipstick urinalys...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The journal of clinical hypertension (Greenwich, Conn.) Conn.), 2024-04, Vol.26 (4), p.349-354
Hauptverfasser: Wang, Liying, Gao, Jinsong, Tang, Pingping, Hu, Huiying, Chen, Xiaoxu, Chen, Ziyi, Sun, Yin
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:To compare the urine Congo-red dot paper test (CRD) with dipstick urinalysis to screen preeclampsia (PE). A total of 409 paired spot urine samples were obtained prospectively from women with suspected pre-eclampsia attending for routine hospital visits. Congo-red dot paper test and dipstick urinalysis were examined and compared to screen pre-eclampsia. The agreement between the two urinary test is modest (kappa coefficient = 0.28, 95% CI 0.14-0.42). The specificity of CRD was higher than urinalysis (97.4% vs. 90.4%, p 