Direct Plasma Deposition of Collagen on 96-Well Polystyrene Plates for Cell Culture

A cold atmospheric plasma unit was used to deposit a biologic, in this case collagen, onto a surface. A collagen coating was applied to 96-well polystyrene plates at a range of powers to determine the effects of the plasma power on the coating structure and viability. Plasma characterization was car...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:ACS omega 2020-10, Vol.5 (39), p.25069-25076
Hauptverfasser: O’Sullivan, Denis, O’Neill, Liam, Bourke, Paula
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A cold atmospheric plasma unit was used to deposit a biologic, in this case collagen, onto a surface. A collagen coating was applied to 96-well polystyrene plates at a range of powers to determine the effects of the plasma power on the coating structure and viability. Plasma characterization was carried out using voltage, current, and power measurements. Coating characterization was completed using gravimetric measurement, cell growth, water contact angle, as well as spectroscopic analysis and compared to commercial collagen-coated plates. Cell culture studies were also undertaken. The plasma coating matched the performance of the commercial plate but dramatically reduced production time and cost. This method could allow for automated inline production of collagen-coated plates for cell culture applications.