Aseptic loosening of an APTIS distal radioulnar joint arthroplasty – Case report
AbstractDysfunction of the distal radioulnar joint can cause significant pain and instability. The self-stabilizing APTIS distal radioulnar joint prosthesis is used as a solution for severe distal radioulnar joint pathologies. We present a case of a 60-year-old male, who received an APTIS distal rad...
Gespeichert in:
Veröffentlicht in: | Trauma case reports 2019-06, Vol.21, p.100196-100196, Article 100196 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | AbstractDysfunction of the distal radioulnar joint can cause significant pain and instability. The self-stabilizing APTIS distal radioulnar joint prosthesis is used as a solution for severe distal radioulnar joint pathologies. We present a case of a 60-year-old male, who received an APTIS distal radioulnar joint prosthesis which resulted in aseptic loosening within five years of the initial implantation. Infection, incorrect implantation, demographic differences and over-activity were all excluded as the source; therefore, mechanical aseptic loosening was concluded. Ultimately, two surgeries were required to resolve the patient's pain, which resulted in a one-bone forearm once the implant was extracted. The solution to a failed APTIS implant, a one bone forearm, is difficult and protracted, so every effort should be attempted to preserve distal ulna bone stock before resorting to the implantation of this device. |
---|---|
ISSN: | 2352-6440 2352-6440 |
DOI: | 10.1016/j.tcr.2019.100196 |