Simplifying biometry in oil-filled eyes: A novel formula for axial length calculation in eyes with 1000 cSt silicone oil

Purpose: To derive a formula for accurate axial length (AL) assessment using routine ultrasound in silicone oil-filled eyes, where optical biometry is unavailable or not possible. Methods: This was a prospective, consecutive, nonrandomized study of 50 eyes of 50 patients conducted in a tertiary care...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Indian journal of ophthalmology 2023-06, Vol.71 (6), p.2466-2468
Hauptverfasser: Kapoor, Monika, Venkatesh, Pradeep, Chawla, Rohan, Temkar, Shreyas, Aggarwal, Esha
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Purpose: To derive a formula for accurate axial length (AL) assessment using routine ultrasound in silicone oil-filled eyes, where optical biometry is unavailable or not possible. Methods: This was a prospective, consecutive, nonrandomized study of 50 eyes of 50 patients conducted in a tertiary care hospital in North India. AL measurements were taken using both manual A-scan and IOL master, both in silicone oil-filled status and 3 weeks after silicone oil removal. A correction factor of 0.7 was used for AL adjustment in oil-filled eyes. The corrected AL (cAL) was compared with IOL master values in oil-filled eyes. Agreement analysis was carried out using Bland Altman plot. Linear regression analysis was done using uncorrected manual AL to find a new equation. Data was analyzed using Stata 14. A P value